[Rinsho Byori 50 : 74-78, 2002]
Application of HPLC Method for the Measurement of Hyaluronan in Pleural Fluid We examined whether high-performance liquid chromatography(HPLC) method
was more suitable for the measurement of hyaluronan(HA) in pleural fluid
than the standardized electrophoretical method. Forty-four samples of pleural
fluid were measured both by the HPLC method and by the electrophoretical
method. The HA values measured by the HPLC method did not show significant
correlation(r0.33) to those measured by the electrophoretical method.
We investigated the reason of this insufficiency of the correlation. The
electrophoretical method was not specific for the measurement of HA and
other glycosaminoglycan, such as chondroitin could be measured by this
method. Therefore, the HA value measured by this method was higher than
that measured by the HPLC method. The other problem was some loss of HA
during the pre-treatment of samples with percloric acid in the electrophoretical
method, which was needless in the HPLC method. We concluded that the HPLC
method was more appropriate than the electrophoretical method for the measurement
of hyaluronan in pleural fluid.
*1Glycoconjugate Research Center, Seikagaku Corp., Yokosuka 239-0831 |