[Rinsho Byori 50 : 226`233, 2002]
Development of Homogeneous Assay for HDL-C in Serum Recent development of several homogeneous assays for measuring HDL-cholesterol(HDL-C) in serum without pretreatment has been reviewed. Several problems on the five homogeneous assays based on PEG-modified enzyme/ż-cyclodextrin sulfate(PEGME), polyanion-polymer/detergent(PPD), immunoinhibition, catalase/detergent and 4-sulfonyl calyx[8]areine are discussed comparing with CDC reference method. The two homogeneous assays(PEGME and PPD) used in many laboratories were precise and correlated well with CDC reference method(mean percentage bias : 5%). The discrepancies of recoveries for apo E-rich HDL-C were observed among reagents of homogeneous assays in serum from patients with hepatobiliary diseases and cholesterol ester transfer protein(CETP) deficiency. Apo E-rich HDL slightly contained in normal serum was most recovered by homogeneous methods, but not by precipitation methods. The differences between the normal values for HDL-C obtained by homogeneous assay(PEGME method) and precipitation method(dextran sulfate-phosphotungstate-MgCl2) were statistically significant(p0.01).
*1Department of Laboratory Medicine, Kumamoto University Hospital, Kumamoto 860-8556 |