We developed a latex turbidometric immunoassay(LTIA) for elastase 1. The
precisions were 1.0`2.8% intra-run and 1.2`2.9% inter-day. The dilution
test showed good linearity between 80`4000ng/dl. There was interference
by bilirubin, hemoglobin, lipemia or rheumatoid factor. Serum and plasma
elastase 1 measured by LTIA( y ) showed significant correlations with those
by RIA( x )(serum : y0.98x|18, r0.980, n64, plasma : y0.96x|19,
r0.976, n61). It was confirmed that the values were elevated in patients
with pancreatic disease compared to the normal level.
These results demonstrate that the determination of elastase 1 in plasma or serum by this method is useful, simple and time-saving, when compared with that by RIA.
[Rinsho Byori 50 : 376`380, 2002]